ENT / Endoscopy Rigid & Flexible repairs


AMP can provide a door to door repair service that fully ISO certified and manufacturer independent for all makes and manufacturers af Rigid scopes, Flexible scopes and Uretherorenoscope's.

Key Service Features

  • All repairs are manufacturer independent
  • All manufacturers and makes can be repaired
  • Repair are all ISO 9001 / ISO 13485 certified
  • 1 year repair warranty
  • Service exchange option if a scope cannot be repaired
  • Loaner scope service / 7-10 working day turnaround

Cover glass defective

Damage:
  • Distal end of the shaft damaged
  • Cover glass slivered
  • Objective system is leaking

Distal end burned by HF-loop

Damage:
  • Fibre optic burned
  • Shaft end burned
  • Leaky

Defective, crushed distal end

Damage:
  • Crushed distal end
  • Broken, leaky fibre optic
  • Defective objective

 


Broken light fibres

Damage:
  • Protrusive cover glass
  • Broken/withdrawn light fibres

Laser damage

Damage:
  • High impact of temperature
  • Causes optic leaks

Shaver touch

Damage:
  • Cover glass or negative scratched
  • Distal shaft tube damaged

 


Broken light fibres

Damage:
  • Twisted shaft
  • Dents in shaft